FTIR Thermo-Nicolet 6700
IR01
CONTACT PERSON:
Dr Fotis Katsaros
EMAIL:
f.katsaros@inn.demokritos.gr
IR01
CONTACT PERSON:
Dr Fotis Katsaros
EMAIL:
f.katsaros@inn.demokritos.gr
FTIR
Tool
Thermoscientific Nicolet 6700
Applications
Characterization of solids, liquids, and gases. Identification of unknown materials and confirmation of production materials (incoming or outgoing). Applications in environment, food, forensics, pharmaceuticals, polymers and plastics, quality control and general analysis.
Technical Specifications
Source: ETC EverGlo Sourcecovering the range 12500-350cm-1
Detectors: 1x DTGS(6000 -370 cm-1), 1x MCT-B (11700-400cm-1)
Beamsplitter: XT-KBr (11000-375cm-1)
Resolution: 0.125 cm-1
Signal to noise ratio: 50000:1
Linearity: 0.07%T
Scan rate: 65 scans/s
Accuracy of wavelength: <0.01cm-1
Additional Tools
ATR cell
DRIFTS cell
High pressure/ Temperature chamber: Tmax:900oC, Pmax: 1500psi, Window materials: ZnSe, Thermocouple: K-type, Cooling: water
Nitrogen purging system