VTI
VTI
CONTACT PERSON:
Dr. Fotios Katsaros
EMAIL:
f.katsaros@inn.demokritos.gr
VTI
CONTACT PERSON:
Dr. Fotios Katsaros
EMAIL:
f.katsaros@inn.demokritos.gr
High Pressure Gas sorption analyzer
Tool
HPVA100SP(Micromeritics former VTI Corp.)
Applications
Gas storage, Gas separations, Study of gas solid interactions
PCT Isotherms
Sorption Kinetics
Adsorption/Desorption Cycling, PCT and Kinetics
Technical Specifications
Static, manometric measurement type
Gases: H2, CO2, CH4, Ethylene, Propane, any permanent gas
Pressure range: vacuum-160bar
Temperature range: 77K–573K
Pressure transducers: 2x vacuum to 200 bar / Accuracy: 0,17 mbar
Outgassing station: Yes,up to 673K
Vacuum: <10-3mbar
Dosing manifold temperature stability: 0.1K
Cell volume range: from 10 ml up to 500 ml
Volume calibration with He
Additional Tools
Cooling system form 243K-323K