DCS C80 calvet

Calorimeter with high precision Calvet detector

Tool

Calvet C80 DSC (Setaram)

Applications

Isothermal calorimeter, Scanning calorimeter, Reaction calorimeter: mixing vessels allow for the simulation of reaction conditions, liquid/liquid, solid/ liquid and even gas phase reactions can be measured, Cp Measurement of liquids, solids and powders, Heats of solution etc.

Technical Specifications

Temperature range: Ambient to 300 °C

Temperature accuracy: +/- 0.1 °C

Temperature precision: +/- 0.05 °C

Programmable temperature scanning rate: 0.001 to 2 °C/min

Enthalpy accuracy: +/- 1 %

Calorimetric Precision: +/- 0.1 %

RMS Noise: 1 μW

Resolution: 0.10 μW

Dynamic range: +/- 660 mW ; +/- 2000 mW

Additional Tools

High pressure cells (up to 100 bar)

Skip to content